Hi Kathereh
I purge the mobile phases and flowmeter and mobile phase is constantly running
through the column, but maybe I should try conditioning the column as you
suggested prior to starting a run.
As for the sonication of the emitter should I submerged it all in the solvent?
If this is the case how do I make sure that the emitter does not float in
solution?
Thanks
Emanuele
-----Original Message-----
From: ABRF Discussion Forum <email obscured>] On Behalf Of Khatereh
Motamedchaboki
Sent: 09 February 2016 21:23
To: <email obscured>
Cc: e.scollo <email obscured>>
Subject: Re: [ABRF Discussion Forum] Bubble forming at the tip of SS emitter
for nano ESI
Hi Emanuele,
Our column and pre-column(trap) equilibration goes up to 500 nl/min with 800 ba
max pressure cap, and we run this for 5-10 min, we also do purge solvents and
flush air to start with, while you are running it at 500 nl/min, isocratic at
50:50 A:B, disconnect the column out to union wait for flow for half min,
connect it again, then disconnect the trap from the union and wait for flow to
flush the union and then connect again, then connect the trap to T connection,
disconnect waste line and column from T, let the let it flow for half min-min,
then connect the column or emitter and let it flow for half min then connect
the waste line, you can sonicate the T union in methanol or isopropanol for 5
min too, rinse it with HPLC grade water then and connect again, make sure there
is no air before connecting the column or emitter, our droplet issue usually
resolves by disconnecting waste line to T while on isocratic flow and then
connect it again, make sure you are tightening it well but not too much.
If this does not resolve the issue ,sonicate the tip, if you are running ug
amount of samples on nano columns you will need to sonicate the tip every 10
days to avoid blocking it fast, we have a thermo tool which we use to hold the
emitter/column tip on top of the beaker to sonicate the tip as I recommended in
my 1st email, The next step would be adjusting the voltage and distance, if all
of this does not help, check your source, it might be the spray source resistor
has gone bad and need to be replaced and it's not applying right voltage in the
tune setting to your column/source.
That has happened to us one, and your service engineer should be able to fix
the source, if the resistor is ok, check it on a new column /emitter then maybe
your column tip or voltage connection is gone bad and you need to replace the
column/emitter, all of this can be done in less than hour procedure.
Please feel free to call me if you have any other questions,
Good Luck,
Khatereh Motamedchaboki, PhD.
Director, Proteomics Facility
Sanford Burnham Prebys Medical Discovery Institute
10901 North Torrey Pines Road
La Jolla, CA 92037
Tel: (858) 646 3100 Ext: 3710
Fax: (858) 795 5221
http://www.sbpdiscovery.org/technology/sr/Pages/LaJolla_Proteomics.aspx
-----Original Message-----
From: ABRF Discussion Forum <email obscured>] On Behalf Of e.scollo
Sent: Tuesday, February 09, 2016 11:39 AM
To: <email obscured>
Subject: Re: [ABRF Discussion Forum] Bubble forming at the tip of SS emitter
for nano ESI
Hi Kathered
Thanks for your reply. I noticed that when I get the emitter very close to the
source then the droplet disappear but then it forms again. Obviously I cannot
have the emitter too close to the source otherwise the spray is not going to be
stable, however I am not sure I understood well how you use the pump script to
get rid of air. The flow on the nano pump is 300 nl/min so do you mean
increasing the flow to remove air?
Also how do I understand whether the issue is the emitter or not?
Thanks
Emanuele
ββ
View topic http://list.abrf.org/r/topic/3XBCw8m8h6lU3hWtMdQ5SU
Leave group <email obscured>?Subject=Unsubscribe
ββ
View topic http://list.abrf.org/r/topic/2IOc0NyIYIjszjHDJGOR30
Leave group <email obscured>?Subject=Unsubscribe
ββ
View topic http://list.abrf.org/r/topic/3G9eKyNr8e01xsgRqaWzpC
Leave group <email obscured>?Subject=Unsubscribe